01 / Start with the task
A guide transfers a defined part of a reviewed plan. Begin by naming that task, the structures supporting the guide and the elements still controlled by the clinician. These are different questions.[14]
An implant guide, a sinus-window template and a corticotomy marking guide transfer different information. Describe the intended job before discussing the design.
Identify where the guide rests. Tooth, mucosal, bone, implant or combined support must be considered in the context of the procedure.
For implant treatment, distinguish pilot guidance, preparation guidance and guidance that also includes implant insertion. Use the terminology of the actual system.
02 / Before CAD
Record selection, interpretation and registration come before guide design. The treating clinician’s assessment and the restorative objective provide the context; a matched dataset does not establish a diagnosis.[2][7]
03 / Support & stability
Support type, distribution and fixation are relevant to accuracy. The systematic review includes both laboratory and clinical evidence; its comparisons do not make one support design appropriate for every case.[13]
Which teeth are used, what is their current condition, and how does the design behave beside a bounded space or a distal extension?
What tissues are represented by the records, what clinical access is intended and how will seating and stability be assessed?
Does the plan need additional stabilisation? Where and how can it be assessed within the actual anatomy and selected system?
04 / Guide + instruments + access
The guidance interface, instruments and clinical access belong in the same discussion as the guide body. The selected components and intended extent of guidance must agree.[14][18]
The figure deliberately separates the parts for explanation. It gives no drilling sequence, sleeve offset, preparation depth or interchangeable-system claim.
05 / The design review
Our teaching uses a design conversation: state the choice, explain its reason, show how it is reviewed and document what needs approval. Use current instructions for the installed software version and the selected system.[15]
Is this a pilot guide, a guide for a defined preparation sequence, implant placement guidance, an access-window template or a marking guide?[14]
Document: State exactly what is guided and what is not.
Which surfaces support the guide? How does the plan address loss of support, tissue movement, a free end or the need for fixation?[13]
Document: Identify the support and the seating-verification strategy.
Do the selected guide interfaces and actual instruments belong to the intended workflow? Can the assembly be used at the planned site?[14][18]
Document: Confirm the system, kit, components and access review.
Can the planned guide be inserted and its seating assessed? How does its coverage relate to the clinical approach and the structures being used for support?[14]
Document: Review the design in context, not only as an isolated mesh.
A guided CAD review / Our teaching framework
01
Identify the reviewed restoration, implant plan or surgical marking. Make clear which objects are reference data and which belong to the proposed guide.
02
Discuss the covered surfaces, support distribution, path of insertion and the clinical approach. Explain how the design can be inspected for seating.
03
Review the selected component information with the actual instruments and intended task. Resolve any uncertainty about the combination before approval.
04
Explain how the guide relates to its support and planned task. A pleasing exterior view should not be the only view used in the discussion.
05
Identify the approved revision and the agreed file, units/scale and companion information. Keep a review copy distinct from a version released for production.
06
Record the remaining questions, who reviews them and what can proceed. If the plan or relevant records change, identify which approval must be revisited.
The result is a reviewed design version with a defined task. A late change to the plan, support, components or clinical records calls for review of the affected design and handover—not an unnoticed change to the exported file.
06 / OsteoPerio design & fabrication
OsteoPerio handles surgical-guide design and fabrication. We discuss the intended use and the complete production route with the requesting clinician, whether the agreed output is a digital design or a printed guide.
Manufacturer workflows connect scanning and design with printing, washing, post-curing, finishing, assembly and preparation for use. These steps are specific to the selected material and equipment; a generic resin setting or cycle is not a substitute for their instructions.[16]
Identify the plan, intended use and final design. Resolve outstanding clinical or technical questions before an export is treated as approved.
Confirm the responsible producer, compatible equipment, intended material and applicable instructions. Agree who supplies and assembles any components.
Follow the selected manufacturing workflow, including the required finishing and post-processing. These steps depend on the actual material and equipment.
Check the produced guide and relevant components, record the released version and resolve discrepancies. A manufacturing check does not replace clinical seating verification.
Confirm identification, instructions, processing state, storage/transport conditions and the receiving clinician’s next checks before dispatch or file delivery.
07 / Different checks / Different questions
Errors can arise at different stages and combine. Review at each handover is therefore part of the workflow, rather than a single final inspection.[18]
Does the guide correspond to the intended model and design? What can be inspected here, and what still depends on the patient and the procedure?
Can the treating team verify the intended seating, support and stability in the actual clinical situation?
Does the planned guide work with the confirmed components and available clinical access? Is the intended preparation understood?
If seating, components, integrity or the current anatomy do not agree with the plan, pause the guided workflow for clinical and technical reassessment. Do not assume pressure or unreviewed alteration will resolve the cause.
Published planned-to-placed deviations describe an accuracy endpoint. They do not guarantee an individual guide’s performance or define a universal case-specific clearance. A clinical team still needs a plan for uncertainty and for an unsuitable guide.[6]
08 / Digital or physical delivery
An STL design and a printed guide are different deliverables. Before work starts, agree what OsteoPerio will provide, what the receiving team will do and who approves the clinical plan.
| Agree in advance | Design + STL | Design + printed guide |
|---|---|---|
| Main output | Reviewed design export and the agreed plan/component information. | The fabricated guide and the agreed components and documentation. |
| Production | Your agreed producer uses the approved export in its manufacturing workflow. | OsteoPerio fabricates the guide under the agreed production arrangement. |
| Processing | Receiving producer/team confirms its required processing and preparation steps. | Supply state and any further preparation are explicitly agreed; shipping does not imply sterility. |
| Before clinical use | Treating team verifies the produced guide, current clinical fit and system compatibility. | Treating team verifies the received guide, current clinical fit and system compatibility. |
| Costs and timing | Confirmed after scope, compatibility and file requirements are reviewed. | Confirmed after scope, production needs, destination and logistics are reviewed. |
This is a discussion framework for agreeing a request, not a claim that every item is supplied in every package. Clinical prescription/approval, design, manufacture and final clinical verification have distinct responsibilities.
09 / Connect the concepts with existing records
Inside a planning case
1 min 24 sec · No narration

The rotating coDiagnostiX view shows the plan for two separate upper-jaw bridges: one supported by three implants and the other by two. This is a partial-arch restoration, not a full-arch bridge. Surface scans, bone anatomy and restorative and guide components help explain the planning relationships.
A digital planning demonstration. The implant number, guide design and treatment stages are specific to the individual case.
Read this example with a design question
Use the existing coDiagnostiX film to identify the restorative groups, the displayed records and the guide components. This is a partial-arch planning demonstration; it does not document guide manufacture, clinical seating or the surgical outcome.

A guide for access-window planning
Follow the existing digital planning film, printed guide and intraoperative sequence. The window guide’s task and implant-placement timing are separate parts of the plan.
Explore the sinus sequenceLinked sequence contains surgical photographs.

A guide for transferring markings
See the model and separate guides, then follow the marking-guide workflow. Surgery: Dr. Sahand Shirzadi Maryan. Guide design and printing: Dr. Behnam Taghavi.
Explore the documented caseThe existing case retains its clinical sequence, credits and early-follow-up limits.
10 / How we structure the teaching
Begin at your current level. We can connect theoretical preparation and article discussion with software-based planning, guide-design review and the production workflow. The modules below show how a complete learning pathway can be organised with OsteoPerio.
Each module has a practical discussion, an output to explain and a checkpoint to defend. Exercises, software access, teaching language, duration and fees are agreed individually. This is a curriculum framework, not a scheduled or accredited course.
Define the treatment objective, proposed restoration and the precise job of the guide. Separate implant guidance, window positioning and surgical marking.[2][14]
Discuss a case brief with the tutor: what will the guide control, what remains under direct clinical control, and which alternatives need consideration?
A written indication, named treating clinician and a list of unresolved questions.
Discussion checkpoint: Can you explain why this guide is needed, rather than beginning with a software feature?
Distinguish radiographic volume, optical surfaces, occlusion, photographs and the restorative proposal. Review their identity, timing, coverage and correspondence.[7]
Inspect a teaching dataset in different views. Explain where the registration is convincing and where you need further information.
A record inventory and a documented registration review.
Discussion checkpoint: A convincing rendered surface is not enough to accept the match.
Relate the intended teeth to anatomy, tissue needs, access and the proposed treatment stages before designing the transfer tool.[2]
Present the plan and an alternative. Identify the assumptions requiring the treating clinician’s review.
A reviewed plan with an explicit version and scope of guidance.
Discussion checkpoint: Guide design follows clinical planning; it does not resolve an uncertain diagnosis.
Compare the support surface, retention or fixation strategy, and the extent of instrument guidance. These are separate design decisions.[13][14]
Compare a bounded space, a free-end situation and an edentulous teaching example. Describe what keeps each proposed guide seated.
A reasoned support concept and system-compatibility list.
Discussion checkpoint: The same total number of implants does not imply the same guide design.
Connect guide coverage, insertion path, inspection access and component placement with the intended clinical task. Use the current software and kit instructions.[15][18]
Walk through a design review with the tutor. Explain how access, support, component interfaces and any proposed changes are checked.
A design-review record and the version submitted for clinical approval.
Discussion checkpoint: Do not substitute a generic thickness, sleeve offset or drilling diagram for the selected system’s instructions.
Trace the approved export through the selected material and production workflow, finishing, component assembly and inspection.[16]
Discuss a production record and the relevant manufacturer instructions. Identify which steps belong to the specific equipment/material combination.
A production and inspection checklist for the agreed workflow.
Discussion checkpoint: A good-looking print does not demonstrate a complete or suitable processing cycle.
Separate model checks, clinical seating, stability, component compatibility and readiness for the intended procedure.[6][18]
Use a teaching guide/model to discuss fit and access. Work through a handover with the design version, component list and remaining clinical checks.
A traceable handover and a plan for resolving any discrepancy.
Discussion checkpoint: Model fit and clinical fit answer different questions.
Connect the guide with clinical judgement, documentation and follow-up. Compare what the plan intended with what was actually observed.[6][14]
Discuss an implant, sinus-window or marking-guide example. Identify what was transferred, what was verified and what the records cannot establish.
A case reflection, an evidence question and a next learning objective.
Discussion checkpoint: Placement accuracy, clinical outcomes and the learner’s competence are different endpoints.
Read the agreed sources and formulate your questions before the session.
Present your reasoning and review a teaching plan or design with the clinical team.
Relate the design to production and, where individually arranged, relevant clinical observation.
For qualified dentists and postgraduate clinicians. Observation does not include operating on patients. The supervised clinical programme remains in preparation and is not available to book. The downloadable brief is an educational discussion aid, not a clinical release form.
11 / Professional guide support / OsteoPerio
You can ask OsteoPerio to plan and design a guide, send an agreed STL design, or fabricate and deliver the printed guide. Start with the task and your clinical setup. We review the request and clarify the appropriate scope before a proposal or production arrangement is confirmed.
Digital delivery
Discuss a guide design for production through your own agreed manufacturing workflow.
Opens a draft in your email app. Review it before sending.
Physical delivery
Discuss guide design and fabrication by OsteoPerio, followed by delivery to the agreed destination.
Opens a draft in your email app. Review it before sending.
Professional review
Start with a focused conversation about your proposed workflow, existing plan or production requirements.
Opens a draft in your email app. Review it before sending.
12 / Questions & critical reading
Yes. We can structure individually arranged teaching from record interpretation and planning foundations through guide design and production review. The starting level, software needs, teaching language, exercises and format are agreed with you. Practical work and observation depend on the arrangement; reading the chapter does not qualify someone to perform surgery.
Yes, you can enquire about planning, guide design and digital delivery for production through your own agreed workflow. We first clarify the clinical task, compatible system, required records, approval and manufacturing responsibilities. The STL is one deliverable; the agreed plan, design version and relevant component information also matter.
OsteoPerio handles guide design and fabrication. Tell us your intended use and delivery country so we can discuss compatibility, scope, documentation, processing responsibilities, costs and logistics. The initial enquiry is a request for review and a proposal, not an accepted production order or guaranteed delivery date.
Ask us to review the proposed file and intended workflow first. A surface file alone does not establish that the clinical plan, design, material/process and components are appropriate. Whether an externally prepared design can be accepted is agreed after review.
No. Software, guide design, materials, components, processing and instruments need to be considered as a connected workflow. Use the instructions applicable to the actual combination rather than transferring numerical settings from an unrelated example.[15][16]
No. It offers a production/design check in a model context. The treating clinician still needs to verify the current clinical situation, seating, stability and access. An unresolved discrepancy requires reassessment before proceeding with that guide.[18]
No. Guide support, surgical access, implant-placement timing and loading are separate decisions. They depend on the clinical plan and the conditions encountered; the existence of a digital guide does not settle them.[2][14]
Start with your professional role, country, the general guide task, the support you need, the system/kit if known and preferred timing. Do not include patient-identifying details or attach clinical files to the initial enquiry. We will clarify the appropriate records-transfer arrangements after discussing the request.
Clinical planned-to-placed deviations and laboratory design comparisons answer different questions. Neither certifies an individual guide or proves long-term benefit.[6][13]
A narrative review and a risk review help organise the system and its possible failure points. They do not validate this chapter as an operative protocol.[14][18]
A manufacturer’s application guide belongs to its own equipment/material context. Regulatory guidance also requires the actual device and production arrangement to be identified.[16][17]
13 / Scientific & technical reading
Scientific sources inform the concepts; manufacturer documents govern their specific systems. The figures, learning exercises and handover framework are original OsteoPerio teaching material. Sources do not certify or endorse OsteoPerio’s production process.
[2] Consensus report / 2018
ITI consensus: digital technologies in implant dentistryWismeijer D, Joda T, Flügge T, et al. Clinical Oral Implants Research. 29(Suppl 16):436–442. doi:10.1111/clr.13309.
Connects digital diagnosis, planning and clinical transfer; digital tools remain adjuncts to comprehensive assessment.
[6] Systematic review and meta-analysis / 2018
Accuracy of static computer-aided implant surgeryTahmaseb A, Wu V, Wismeijer D, Coucke W, Evans C. Clinical Oral Implants Research. 29(Suppl 16):416–435. doi:10.1111/clr.13346.
Clinical studies report deviations between planned and placed implants. Mean values do not define a guaranteed result or a universal case-specific clearance.
[7] Registration accuracy study / 2017
Registration of CBCT data and intraoral surface scansFlügge T, Derksen W, Te Poel J, et al. Clinical Oral Implants Research. 28(9):1113–1118. doi:10.1111/clr.12925.
Thirty-six patient datasets; examines reconstruction, operator and restoration-related influences. These system-specific findings do not certify every planning workflow.
[13] Systematic review / 2023
Accuracy of digital surgical guides: support, design and manufactureShi Y, Wang J, Ma C, Shen J, Dong X, Lin D. International Journal of Implant Dentistry. 9:38. doi:10.1186/s40729-023-00507-w.
Includes laboratory and clinical studies. Design comparisons and manufacturing findings need to be read within those different settings.
[14] Narrative review / 2025
Static computer-assisted implant surgery: terminology, systems and limitationsYeo XH, Fan S, Chantler JGM, Chow J, Pimkhaokham A, Mattheos N. Australian Dental Journal. 70(Suppl 1):S129–S145. doi:10.1111/adj.70007.
Explains guide and instrument-system concepts. An overview is not a validated recipe for every design or clinical indication.
[15] Manufacturer instructions portal / 2026
coDiagnostiX: current instructions for useDental Wings. Official electronic instructions portal. Accessed 30 October 2026; select the applicable version and language.
Use the instructions relevant to the installed software. Listing this resource does not establish the software version used for a particular case.
[16] System-specific technical guidance / 2026
A manufacturer example: scan-to-production surgical-guide workflowFormlabs Dental. Application Guide: 3D Printing Surgical Guides. Official resource, accessed 30 October 2026.
An example of connected manufacturing steps. Its settings are not universal or a statement of OsteoPerio’s equipment, material or processing protocol.
[17] Regulatory guidance / 2021
Custom-made, adaptable and patient-matched devices: MDCG 2021-3Medical Device Coordination Group. Questions and Answers on Custom-Made Devices. March 2021. Non-binding regulatory guidance.
Distinguishes device categories and responsibilities. A patient-specific design is not automatically a custom-made device or proof of compliance.
[18] Clinical risk review / 2019
Guided implant surgery risks and their preventionTatakis DN, Chien HH, Parashis AO. Periodontology 2000. 81(1):194–208. doi:10.1111/prd.12292.
Discusses errors across the workflow, the need for verification and the importance of system knowledge and surgical experience.