[1] Registration accuracy study · 2017
Registration of cone beam computed tomography data and intraoral surface scans — A prerequisite for guided implant surgery with CAD/CAM drilling guides (new tab)Flügge T, Derksen W, Te Poel J, et al. Clinical Oral Implants Research. 28(9):1113–1118. doi:10.1111/clr.12925.
Full text verified. Thirty-six patient datasets were registered by four examiners. Reconstruction settings, user and restoration-related factors affected measured landmark deviations. The reference and commercial workflow constrain interpretation; this is neither a universal numerical acceptance rule nor proof that a screen overlay is ground truth.
[2] Clinical accuracy study · 2021
Comparison of Dental Surface Image Registration and Fiducial Marker Registration: An In Vivo Accuracy Study of Static Computer-Assisted Implant Surgery (new tab)Han YT, Lin WC, Fan FY, et al. Journal of Clinical Medicine. 10(18):4183. doi:10.3390/jcm10184183.
Full text verified. Thirty patients/33 surgeries/74 implants; the principal fully guided comparison included 56 implants. Surface versus fiducial registration was chosen according to case conditions, not randomized. Desktop scans of stone casts supplied the optical data. The clinical endpoint was planned-to-placed implant deviation, not isolated registration error.
[3] Retrospective registration reprocessing study · 2022
Evaluation of different registration methods and dental restorations on the registration duration and accuracy of cone beam computed tomography data and intraoral scans: a retrospective clinical study (new tab)Piao X-Y, Park J-M, Kim H, Kim Y, Shim J-S. Clinical Oral Investigations. 26(9):5763–5771. doi:10.1007/s00784-022-04533-7.
Abstract verified; full text not available for this review. Thirty-six jaws were reprocessed with five software methods. Registration duration differed; accuracy differences between methods were not statistically significant. Restoration proportion was associated with accuracy. This does not establish equivalence or a preferred method for every case.
[4] Patient-dataset segmentation/registration study · 2026
Effect of the CBCT data segmentation threshold on registration accuracy with surface scanning (new tab)Lo Russo L, Lorusso M, Ercoli C, Sorrentino R, Ciavarella D, Ruggiero G, Esperouz F. Journal of Prosthetic Dentistry. 135(3):595.e1–595.e10. doi:10.1016/j.prosdent.2025.10.044.
Abstract verified; full text not available for this review. Nineteen patients, seven segmentation thresholds and six reference points. Segmentation influenced measured registration deviations, particularly in the apicocoronal direction. These observations do not identify a universal threshold setting or directly measure subsequent clinical implant-placement outcomes.
[5] Systematic review and meta-analysis · 2018
Accuracy of static computer-aided implant surgery (new tab)Tahmaseb A, Wu V, Wismeijer D, Coucke W, Evans C. Clinical Oral Implants Research. 29(Suppl 16):416–435. doi:10.1111/clr.13346.
Systematic review and meta-analysis of clinical static-guided implant surgery. Its endpoint is deviation between planned and placed implants across the complete workflow. Pooled means do not isolate registration error, predict an individual result or supply a universal anatomical clearance. Abstract verified.
[6] Consensus report · 2018
ITI consensus: digital technologies in implant dentistry (new tab)Wismeijer D, Joda T, Flügge T, et al. Clinical Oral Implants Research. 29(Suppl 16):436–442. doi:10.1111/clr.13309.
Consensus report linking digital records, planning and clinical transfer. Comprehensive assessment and workflow knowledge remain necessary. It is not a device-specific registration instruction or a universal numerical release criterion. Abstract/consensus record verified.